The 95 – 135 litre toploading autoclave range offers increased chamber sizes compared to the ‘Compact’ units, as well as featuring advanced options not available on smaller sized autoclaves, including Heated Jackets, Integral Steam Generators, Simple/Advanced Vacuum and Cooling options.
These options enable a top loading sterilizer to be tailored to the needs of the customer by speeding up cooling, dealing with liquids, porous-type loads and allowing a wider range of items to be efficiently processed.
Power requirements:
The power requirements for the standard machines are listed in the Specifications table, however these can vary depending on the options selected. Options that affect the power requirement are Integral Steam Generator and Advanced and Simple Vacuum.
For details on the exact power requirements on these options please contact us. N.B. A Neutral line and protective Earth are required for all electrically heated units.
Water and drainage requirements:
A cold water supply of 2-6 Bar minimum, 4 litres/min is required for the ‘Autofill’, vacuum and water cooling options. Max temperature 25°C, Max flow rate 20 litres/min. Requirements vary for RO/de-ionised/hard water. Drainage: Free venting, non-manifolded drain (35mm diameter) capable of withstanding temperatures up to 100°C.