The PSL Rheotek RPV-1 Pulp & Cellulose viscometer system automatically measures the viscosity of a solution of pulp, paper or cellulose dissolved in cupriethylenediamine (also known as CED or Cuene). The RPV-1 Pulp & Cellulose viscometer automates the manual steps set out in standard methods, minimizing the variability of results.
Systems can be configured with 1 to 4 measuring positions. Other options include sample preparation, sample agitation and auto sampler.
Flow times are measured automatically using ASTM Ubbelohde viscometer tubes supplied with ISO 17025 certificates of calibration.
Cleaning is carried out in-situ utilizing solvents and a safe vacuum system.
The comprehensive RPV-1 software program records all measurement data as well as calculating a wide range of results including relative viscosity, intrinsic viscosity (Limiting Viscosity Number), degree of polymerization (DP), K value, kinematic viscosity (mm²/s) and dynamic viscosity (mPa.s).
PSL Viscometer tube types: AKV Ubbelohde or AKV Cannon Fenske Routine
Test Methods Pulp and Cellulose: |
SCAN-CM 15:99 Viscosity in Cupriethylenediamine Solution |
ISO 5351 Determination of limiting Viscosity Number in Cupriethylenediamine (CED) Solution |
ASTM D1795 Standard Test Method for Intrinsic Viscosity of Cellulose |
TAPPI T 230 Viscosity of Pulp (capillary viscometer method) |
PACTAC G.24P Cupriethylenediamine Viscosity of Pulp |
Other Test Methods: |
ASTM D445 Standard Test Method for determination of kinematic viscosity |
ISO 3104 Petroleum products – determination of kinematic viscosity |
IP 71 Section 1 Kinematic Viscosity |
Let's get to know each other
Would you like to purchase one of our products, do you have any question about a product or about services or do you need further information?
We are always just one phone call away or you can send us a message here