Proudct Purpose
Odor test chamber can be used to test and confirm the odor detection and testing with small sampling bag method of storage and usage of such products as electrician,electronics,automotive electrical materials,agricultural products and leather under high temperature.
Feature
Accurate control of temperature and wind speed is achieved through advanced controller,among which air circulation system within the chamber controls the direction and flow rate of air flow in the chamber,while program settings of the control system reaches the function of temperature change and provides the environmental conditions required for odor test automatically.High temperature chamber is sealed with low odor volatile materials or full welded,and its air should be exchanged with clean air outside.
Reference Standards
Model |
V-0V-216 |
V-0V-768 |
V-0V-1000 |
V-0V-4000 |
V-0V-8000 |
|
Volume (L) |
216 |
768 |
1000 |
4000 |
8000 |
|
Performance |
Temperature range |
Environmental temperature ±20℃ ~ +200℃ |
||||
Temperature fluctuation |
≦100℃, 0.4℃; ≦200℃, 0.8℃ |
|||||
Temperature deviation |
≦100℃, ±1.5℃; ≦200℃, ±2.0℃ |
|||||
Heating-up time |
Environmental temperature ±20℃ ~ +200℃ |
|||||
≦40min |
≦60min |
≦40min |
≦60min |
|||
Dimension of the inside(mm) |
W600×H600×D600 |
W830×H1200×D800 |
W1000×H1000×D1000 |
W2000×H1000×D2000 |
W2000×H2000×D2000 |
|
Dimension of the outside(mm) |
W800×H1200×D1295 |
W1030×H1800×D1495 |
W1730×H1480×D1275 |
W2700×H1800×D2750 |
W2700×H2400×D3200 |
|
Temperature and humidity regulating method |
Forced hot wind circulation Ventilation mode |
|||||
Working environment temperature |
(0 ~ +40)℃ |
|||||
Material |
Shell |
Double galvanized plates ,surface spray treatment |
||||
Inner tanks |
Stainless steel SUS201/8K ,Mirror plate |
|||||
Heat perservation |
Glass fibre |
|||||
Heater |
Ni Cr alloy fin type electric heater |
|||||
Draught fan |
Axial flow fan |
|||||
Ventilation door |
Manual switching(opening degree is adjustable) |
|||||
Controller |
Operation mode |
Program operation controller |
||||
Setting range |
0℃ ~ +210℃ |
|||||
Setting accuracy |
0.1℃, 1MIN |
|||||
Display resolution |
0.1℃, 1MIN |
|||||
Input |
K type thermocouple |
|||||
Operating interface |
According to the setting mode |
|||||
Controlling method |
PID+SSR |
|||||
Auxiliary function |
Input disconnection detection detection function, upper limit temperature alarm function, buzzer alarm funtion, automatic temperature overheat prevention function,abnormal indication, alarm number display, self-diagnosis function and temperature and time keeping function. |
|||||
Installed capacity(kw) |
6 |
8 |
9 |
15 |
20 |
|
Power |
AC380±10%V 50Hz 3 phases 4 lines+protective earth wire |
|||||
Standard configuration |
2 sets of sample holders, 1 piece of power line(6 meters long)and 1 set of rubber plug and cap of pin hole |
|||||
Mounting devices |
Leakage circuit breaker for the power, Overheat protector(Upper limit temperature alarm function, fan motor overheat protection swith, control circuit short circuit protection fuse and over temperature protector) |
|||||
Optional |
Pin hole(Φ50 /Φ100mm)
|
|||||
Additional products for optional choice
|
Standard odor smell chamber can be provided, uniform smell test can be realized, and accurate control of temperature, relative humility, wind speed and indoor air cleanliness can be provided. External standard gases can be controlled with micro elements such as solenoid valve(handle valve) and flowmeter, so that the result of smell test can be more stable, accurate, scientific and standardized. |
Let's get to know each other
Would you like to purchase one of our products, do you have any question about a product or about services or do you need further information?
We are always just one phone call away or you can send us a message here